Helpful Hints for Using Exalpha Immunofluorescence Kits

Helpful Hints for Using Exalpha Immunofluorescence Kits Issue & Tips Autofluorescence One method of checking for autofluorescence is to look at the slide with multiple filters. If the same signal is seen with multiple filters, it could be autofluorescence. To help decrease autofluorescence, add a Sudan Black step after the SA-FL incubation. Make up the Sudan Black at 0.2% in … Continue reading Helpful Hints for Using Exalpha Immunofluorescence Kits